1. at full load, 100 ~ 240 Vac input, 20 MHz bandwidth oscilloscope, each output terminated with a 10 µF aluminum electrolytic and 0.1 µF ceramic capacitors.
PART NUMBER KEY
ETMA050400UD
-
XX
-
IC
-
CXX
Base Number
Reserved for Custom
Configurations
example of 5 Vdc, 4 A
DC Plug Type:
See output plug options
on page 3 for more details
Factory
Designation
cui.com
For more information, please visit the
product page
|
For information on related products, view
Dc Power Jacks
CUI Inc
│
SERIES:
ETMA 30W
│
DESCRIPTION:
MEDICAL AC-DC POWER SUPPLY
date
07/10/2015
│
page
2 of 4
INPUT
parameter
voltage
frequency
input current
inrush current
no load power consumption
at 240 Vac
conditions/description
min
90
47
typ
max
264
63
0.6
50
0.3
units
Vac
Hz
A
A
W
OUTPUT
parameter
line regulation
1
load regulation
2
voltage accuracy
hold-up time
switching frequency
temperature coefficient
Note:
1. measured from 100 ~ 240 Vac, full load
2. measured from 60% to full load and from 60 ~ 20% load (60% ±40% load)
conditions/description
5 Vdc output
9 Vdc output
12, 15 Vdc outputs
18, 24 Vdc outputs
at 115 Vac
min
typ
±1
±6
±4
±3
±2
±2
16
70
±0.05
max
units
%
%
%
%
%
%
ms
kHz
%/°C
PROTECTIONS
parameter
over voltage protection
short circuit protection
conditions/description
TVS component to clamp
continuous, auto restart
SAFETY & COMPLIANCE
parameter
isolation voltage
safety approvals
EMI/EMC
MTBF
RoHS
conditions/description
input to output
IEC 60601-1, EN 60601-1, UL 60601-1
EN 55011, EN 60601-1-2, EN 61000-3-(2,3)
as per MIL-HDBK-217F, 115 Vac, 25 °C
2011/65/EU
300,000
hours
min
typ
max
5,656
units
Vdc
ENVIRONMENTAL
parameter
operating temperature
storage temperature
humidity
non-condensing
conditions/description
min
0
-20
typ
max
40
85
93
units
°C
°C
%
cui.com
For more information, please visit the
product page
|
For information on related products, view
Dc Power Jacks
CUI Inc
│
SERIES:
ETMA 30W
│
DESCRIPTION:
MEDICAL AC-DC POWER SUPPLY
date
07/10/2015
│
page
3 of 4
MECHANICAL
parameter
dimensions
ø3.5 +0/-0.1
conditions/description
110 x 50 x 30 (4.331 x 1.969 x 1.181 inch)
IEC320 / C8
min
typ
max
units
mm
input plug
weight
14±0.5
220
g
MECHANICAL DRAWING
positive
*Tip
units: mm
P1R
50 0.5
C
A
B
30 0.5
Right Angle PXR
110 0.5
LED
A
Right Angle PXXR
B
C
DC OUTPUT PLUG OPTIONS / DC CORD
D
cord length
25±5
9±2
Standard PX
A
P5/P5R
P6/P6R
5.5
5.5
B
2.1
2.5
Right Angle PXR
C
9.5
9.5
Unit
mm
mm
MODEL NO.
ETMA050400UD
ETMA090300UD
ETMA120250UD
ETMA150200UD
CABLE GAUGE
18 AWG
18 AWG
18 AWG
18 AWG
18 AWG
18 AWG
CORD LENGTH
720 mm ±50
1,220 mm ±50
1,800 mm ±50
1,800 mm ±50
1,800 mm ±50
1,800 mm ±50
PXX
X X
Plug Type
Plug Angle:
“Blank” = Standard
R = Right Angle
Plug Polarity:
“Blank” = N/A
P = Center Positive
N = Center Negative
ETMA180165UD
ETMA240125UD
act CUI for additional output plug options.
cui.com
For more information, please visit the
product page
|
For information on related products, view
Dc Power Jacks
CUI Inc
│
SERIES:
ETMA 30W
│
DESCRIPTION:
MEDICAL AC-DC POWER SUPPLY
date
07/10/2015
│
page
4 of 4
REVISION HISTORY
rev.
1.0
1.01
1.02
1.03
description
initial release
updated P7/P7R B dimension
V-Infinity branding removed
updated datasheet
date
12/02/2011
03/23/2012
08/21/2012
07/10/2015
The revision history provided is for informational purposes only and is believed to be accurate.
Headquarters
20050 SW 112th Ave.
Tualatin, OR 97062
800.275.4899
Fax 503.612.2383
cui.com
techsupport@cui.com
This device complies with part 15 of the FCC Rules. Operation is subject to the following two conditions:
(1) This device may not cause harmful interference, and
(2) this device must accept any interference received, including interference that may cause undesired operation.
CUI offers a one (1) year limited warranty. Complete warranty information is listed on our website.
CUI reserves the right to make changes to the product at any time without notice. Information provided by CUI is believed to be accurate and reliable. However, no responsibility is
assumed by CUI for its use, nor for any infringements of patents or other rights of third parties which may result from its use.
CUI products are not authorized or warranted for use as critical components in equipment that requires an extremely high level of reliability. A critical
component is any component of a life support device or system whose failure to perform can be reasonably expected to cause the failure of the life support device or system, or to